Shared on behalf of Quality and Risk Management
Please share with applicable staff.
Thunder Bay Regional Health Sciences Centre has been alerted to issues with the Baxter Clearlink System Continu-Flo Solution Set / IV Tubing (Product Code 2C8519) where there is a potential for leaks that stem from faulty upper or lower ports/valves/claves during drug preparation or administration.
The following mitigation actions have been identified by Ontario Health (Cancer Care Ontario):
- Remove lots R25C29185, R25C25161, R25C13147, R25C17079, R25C20130, using
alternate product lots. - Check lines carefully while diluent runs prior to drug reaching the port.
- Document IV tubing lot number and expiration date used for all infusions.
- Monitor and periodically check port/valve/clave and connection sites during changes in
infusion rates.
If you notice any leaking while using any lots of the Clearlink System Continu-Flo Solution Set
/IV Tubing, please take the following actions:
- Immediately stop the infusion and replace the tubing.
- Complete an Incident Learning System and reference the lot and expiration date of the solution set / IV tubing used.
- Notify your Manager and provide him/her with the faulty product.
We appreciate your efforts and cooperation in these extra steps to ensure patient safety.